Products for Industry
Inert Oven DN611I
Item Code : 212904
Model : DN611I
Inert Oven (DN611I )
Product Summary
Suitable for Low Oxidation Environment
Features of Inert Oven (DN611I )
- Suitable for curing process in a low oxygen atmosphere.
- Suitable for heat insulation test and curing process up to 360°C.
- Features
- Simple operation by interactive key input
- Features
- Operation monitor visualizes controller status, temp and temp. changing.
- Features
- Incorporates with maximum 99 steps, 99 patterns program controller with repeat function.
- Features
- Loaded with total operation hours timer.
- Features
- N2 gas flow amount controllable.
Specifications
Specifications
Item Code | 212904 |
---|---|
Model | DN611I |
Method | Forced Convection |
Operating temp. range | Room Temp. +15 to 360°C |
Temp. adjustment accuracy | ±0.2℃(at 360℃) |
Temp. fluctuation | 1.2℃(at 360℃) |
Temp. uniformity | ±3℃(at 360℃) |
Temp. gradient | 20℃(at 360℃) |
Max. temp. reaching time | Approx. 60 min. |
Nitrogen substitution time required | Approx. 70 min. (Ordinary temp with Oxygen concentration of 2%) |
Interior | Stainless steel plate |
Exterior | Cold rolled steel plate with baked melamine resin coating |
Heat insulator | Glass wool + Ceramic fiber |
Heater | Stainless pipe heater 4.0kW |
Sensor | K thermocouple for temperature control and independent overload prevention device |
Fan type / Motor | Sirocco Fan / Condenser type |
Flow meter, Gas carrier | Max. flow 30L/min, O.D. 9mm hose nipple |
Temp. controller | PID control by microcomputer |
Temp. display type | Temp. display: Digital display by 4 digit green LED (resolution:1oC) Setting temp. display: Digital display by 5 digit orange LED (resolution:1oC) |
Timer / Timer resolution | 1min. to 99hrs. 59mins. / 1min. |
Operation function | Fixed temp. operation, Auto-start, Auto-stop, Quick autostop, Program operation |
Program mode | Repeatable operation function up to max 99 steps or 99 patterns. |
Additional functions | Power on and operation time integrating function (up to 65535 hours), Calendar time (24 hours), Calibration offset, Monitor display of integrated power consumption |
Additional functions | Total CO2 emissions and heater operating output, Power failure recovery mode, Save and read out of user settings |
Heater circuit control | Triac with Zero-cross |
Safety device | Self diagnostic functions (Sensor failure, SSR short circuit, Heater line disconnection, Main relay contact damaged, Automatic overheat prevention), |
Safety device | Key lock function, Independent overheating prevention, Electric leakage breaker, Door switch |
Shelf max. load | Approx. 30kg / shelf |
Shelf support step number/pitch | 17pcs. / 30mm |
Shelf plate / support | Stainless wire, 2 pcs. / 4 pcs |
Internal dimensions(W×D×H mm) | 620×600×600mm |
External dimensions(W×D×H mm) | 790×845×1065mm |
Remarks(dimensions) | 223L |
Power Source |
AC220V
20A
Single-phase |
Weight | 130kg |
Maintenance and warranty
Conditions: temperature and humidity: 23 deg.C+/-5 deg.C, 65%RH+/-20% (no load)
External dimensions do not include protrusions.