Products for Industry
Vacuum Drying Oven DP610P
Item Code : 212160
Model : DP610P
Vacuum Drying Oven (DP610P)
Product Summary
Vacuum pump linkage automatic sequence type which release the manual operation of a vacuum pump and a leak valve.
Features of Vacuum Drying Oven (DP610P)
- Automatic sequence vacuum drying oven
- Don't have to worry about the incorrect operation, reverse flow of vacuum pump oil when power off.
- Features
- Automatic sequence operation of vacuum pump linkage is possible.
- Features
- Saving space, high quality controller, individual overheating prevention and many others function.
- Features
- Prevent the reverse flow of vacuum pump oil when power off.
- Features
- There is no exhaust loss because of 1 inch solenoid vacuum valve.
- Features
- Automatic/manual switch is set up.
Specifications
Specifications
| Item Code | 212160 | 
|---|---|
| Model | DP610P | 
| Method | Vacuum drying by decompressed chamber direct heating | 
| Operating pressure range | 101 to 0.1 kPa (760 to 1 Torr) at absolute pressure | 
| Max. temp. reaching time | Approx. 100 min. | 
| Interior/Exterior material | Stainless steel/Cold rolled steel plate with baked-on melamine resin finish | 
| Heater | Mica heater, 3.15 kW | 
| Observation window | Tempered glass and polycarbonate resin plate | 
| Vacuum gauge | Bourdon tube type, 0 to 0.1 MPa (Gauge pressure) | 
| Temp. control method | PID control by microprocessor | 
| Additional functions | Calendar timer (max. 24 Hrs.), Integration time (max 65535 Hrs.), Time display, calibration offset, power discharge, CO2 discharge, heater consumption display, power failure stagnation mode, user configuration information | 
| Heater circuit control | Triac zero-cross control | 
| Operating temperature range | 40~200℃ | 
| Temp. adjustment accuracy | ± 1.5°C (at 200°C) | 
| Operation functions | Fixed temperature operation, auto-start operation, auto-stop operation, program operation ( 99 steps, 99 patterns repeat operation) | 
| Temp. sensor | K-thermocouple (double sensor) | 
| Safety countermeasures | Self diagnosis functions(Sensor, Heater, Triac, automatic overheating prevention), Independent overheating prevention, Key lock function, Electric leakage breaker | 
| Shelf support step number/pitch | 4 steps/140mm | 
| Exhaust port / Purge port | NW25 flange / Rc 1/4 | 
| Accessories Shelf plate | Punched stainless steel, 2 pcs | 
| Internal dimensions(W×D×H mm) | 600×600×600mm | 
| External dimensions(W×D×H mm) | 820×819×1650mm | 
| Remarks(dimensions) | 216L | 
| Power Source | AC220V
																																					16A Single-phase | 
| Weight | 310kg | 
Maintenance and warranty
Conditions: temperature and humidity: 23 deg.C+/-5 deg.C, 65%RH+/-20% (no load)
External dimensions do not include protrusions.