Products for Industry
Cold Trap (CA801)
Product Summary
Efficiently traps water vapor and toxic substances discharged from vacuum oven and rotary evaporator to protect the vacuum pump.
Features of Cold Trap (CA801)
- Features
- Excellent choice to extract acid and organic solvents with the optional glass condenser.
- Features
- Efficiently reduces vapor inhalation amount to the vacuum pump.
- Features
- Can be used as a low temperature tank as well as pre-cooling tank.
- Features
- Utilizes R404A (CA301) and R600a and mixed coolants (CA801)
- Features
- Space saving and highly mobile on wheels
Specifications
Specifications
Item Code | 221437 |
---|---|
Model | CA801 |
Accessories | Casters fixing holder×4 |
Method | Direct trap or Glass trap (optional) |
Dehumidifying capacity | Max. 1.0kg(Water type liquid) |
Max. low temperature | -85℃ |
Refrigerator | Air cooling, 350W |
Refrigerant | R600a and other mixed coolant |
Cooling coil | Installed at tank periphery |
Lid | SUS304, with nozzle (OD ø17.6mm) |
Bath shape / material | Cylindrical / SUS304 |
Ambient temp. range | 5 to 30℃ |
Temperature display | 7 segment LED |
Temperature sensor | Platinum resistance temperature detector Pt100Ω |
Defrosting mechanism | Hot gas bypass system |
Safety device | Electric leakage breaker with over current protection, Refrigeration overload relay, Refrigerator delay timer, Refrigerator high voltage error, Sensor disconnection error |
Sink dimensions | I.D.153×H235mm |
External dimensions(W×D×H mm) | 405×500×850mm |
Remarks(dimensions) | Internal capacity : Approx. 4L (Liquid 3L) |
Power Source |
AC115V/AC220V
Single phase with step-down transformer |
Weight | 65kg |
Maintenance and warranty
Conditions: temperature and humidity: 23 deg.C+/-5 deg.C, 65%RH+/-20% (no load)
External dimensions do not include protrusions.