SINCE 1889 Yamato Scientific Co.,Ltd.
*
Spray Dryer
ADL310
*

The ADL310 is a compact spray dryer using a spray dry method which enables simple use experiments.

Processing Capacity
50g to 300g
Operating Temperature Range
40deg.C to 200deg.C
Liquid Sample Flux
up to 28 ml/min..
Spraying Nozzle
for Liquid and Air

PDF Catalog
  • No danger of damaging foodstuffs, medical products and products that are sensitive to biochemical-heat.
  • No oxidation takes place and the water content of the formed powders is low, so that there is no contamination.
  • Even a solid body of ca.0.5g can be dried.

Specifications
  ADL310
Performance
Water evaporation rate Max. approx. 1,300 ml/h
Temperature control range 40 to 200deg.C
Temperature control accuracy ±1deg.C
Dry air flow rate Max. 0.7 m3/min.
Flow rate for Sample liquid feed pump Max. 28 ml/min.
Structure
Temperature controller PID Digital temp. controller
Heater 2 kW
Aspirator Bypass-type blower
Sample liquid feed pump Quantitative peristaltic pump
Spray pressure gauge Bourdon tube pressure gauge, measurement range : 0 to 0.3MPa (0 to 3kg/cm2)
Spray nozzle Two-liquid nozzle, 1A
Safety Device
  Self diagnosis functions (Temp. abnormal), Sample feed reversion system, Nozzle orifice cleaning(Pulse jet system, Manual)
Standard
Power source (50/60Hz) AC 200/220V, single phase
16.5A/15A
External dimensions(W×D×Hmm) 580×420×1,075
Weight Approx. 78kg
Accessories
Sample liquid tube Silicone tube 2 pcs.
Optional Accessories
Drying air flow meter / Temp. recorder for inlet and outlet temp.

Control panel  
*
* Inlet temperature, outlet temperature are displayed digitally. Condition setting dials and stage level control switch are located on the front panel for easy use.
ADL Control panel
* In this Spray Dryer, the sample are dried and transformed into a uniform fine powder of amorphous form, so both its solubility and reactivity are extremely high in comparison to crystal.
  *

Easy operation and maintenance
ADL Easy operation and maintenance

 

* Inlet temperature, outlet temperature and drying air volume are displayed digitally. Condition setting dials and stage level control switch are located on the front panel for easy use.

* Drying chamber, cyclone, product vessel can be easily detached and cleaned.

ADL Easy operation and maintenance


Spraying nozzle
Nozzle size (μm)
Model 1A 1S 2A 2S 3
Nozzle No. 1650 2050 2050 2850 2850
(F) A 406 508 508 711 711
B 1270 1270 1270 1270 1270
Nozzle No. 64 64 70 70 64
(A) C 1626 1626 1778 1778 1626
Liquid nozzle (F)
ADL310 Liquid nozzle
Air nozzle (A)
ADL310 Air nozzle

*System diagram
*